24 Life Technologies
™
| Incubator Parameters
When using a shaking platform inside of a static incubator, be sure
to temperature-map the platform so as not to incubate cells over hot
spots (and to ensure that you achieve the desired culture temperature).
Don't use a bacterial culture platform (smaller than ¾" in diameter) for
mammalian cell culture.
What incubator parameters do you use for mammalian cell culture?
Table 2 provides the incubator parameters used at Thermo Fisher
Scientific with an Infors Multitron Incubator with a 25 mm throw.
If your shaking incubator has a different throw, we recommend
that, prior to initiating cell line development, CHO-S
®
cell growth be
optimized by adjusting shaker speed based on the following equation:
RCF = 1.118 x 10
-5
x throw x rpm
2
.
Parameter Value
Temperature 37°C
CO
2
concentration 8%
Relative humidity 80%
Agitation 150 rpm (shake flasks)
130 rpm (shaking well plates)
Platform size 850 x 470 mm
Throw 25 mm
Incubator Parameters
Q:
A:
Table 2. Incubator parameters.
Don't:
Do: